news information INDEX > news information > News Center > Calibration and parameter setting of electronic hook scale

Calibration and parameter setting of electronic hook scale

Time:2020.06.24     News sources: Hebei Chaoda Sealing Products Co., Ltd

In the process of self-check of electronic hook scale, press [2] all the time to enter the calibration and parameter setting state. Press [2] to switch between calibration (display CALSPE), parameter setting (display SEt--), internal code display (display - AD -) and instrument interchange (display CHAnGE). Press [1] to access the required function. Press [4] to enter the weighing state under the switching state.

1, the calibration

When displaying CALSPE, press [5] to enter calibration and display CAL000 to indicate zero calibration. After confirming the balance and stabilizing lamp, press [5] to complete the zero calibration. Display CAL001 to prompt the user to start the calibration of load point one, at which point the user enters the weight value of known load one. (Input method: Press [1] to change the number from 0 to 9. Press [2] to shift the number forward when the number is needed.) When the load is on and the stability light is on, press [5]. If the weight value entered at this time is within the normal range, CAL002 will be displayed. At this point, press [4] to complete the linear calibration. If nonlinear calibration is to be carried out, load point two should be calibrated in the same way as the load point. Similarly, the load point three can also be calibrated. However, in addition to zero point, only three load points can be calibrated at most. When the load point three calibration is completed, press [5] and the input weight value is normal or the user presses [4] at any time during the calibration, the program will exit the calibration process and enter the weighing state.

2. Parameter setting

When displaying SEt--, press [5] to enter parameter setting and display D XXX, in which XXX represents the current dividing value. Press [2], then the indexing value cycles between 0.0.0.0.5 1.0 2.0 5.0 10.0 20.0 50.0 0.1. When switching to the required indexing value, press [5] to complete the setting of the new indexing value and switch to the filtering mode. FiLt X is displayed, where when X=0, stable filtering is adopted. When X=1, fast response filtering is adopted. You can switch between the two filtering modes by pressing the [2] key. When switching to the required filtering mode, press [5] to complete the setting of the filtering mode. And go to range Settings to display RXXXXX. Where XXXXX is the current range value. Press [2], then the range will switch between 600, 1000, 1500, 2000, 3000, 5000, 10000, 15000, 15000, 20000, 30000, 50000, 60000. When switching to the required range, press the [5] key to complete the setting of the new range value and enter the setting of the boot setting range to display SEtxxx, where XXX is the current boot setting range. Press [2], then switch between 0, 2, 5, 10, 20, 50 and 100. These values represent the range of percentages. Press [5] to complete the setting of the new power-on zero range, and switch to the setting of accumulation mode. Aut -- X, x=0 means non-automatic accumulation, x=1 means automatic accumulation. Press [2] to switch to the accumulation mode. Press [5] to complete the new accumulation mode setting and enter the zero tracking setting. When the zero trace is set, L0 x is displayed, where x is the different speeds of the zero trace. When x is 0, zero trace is off. When X is 1, it means slow zero tracking, when X is 2, it means medium zero tracking, and when x is 3, it means fast zero tracking. You can switch between the above four modes by pressing the [2] key. When switching to the desired mode, press [5] to complete the setting of zero tracking and enter the setting of time. When setting the time, display TTXXXX, where XXXX is the minute and second when entering the time setting. Use [1] and [2] to enter the new 6-bit time (hour, minute and second). Press [5] to complete the new time setting if the input time is valid. Returns function toggle status.

3. A/D internal code display

When displaying - AD -, press [5] key to display XXXXXX, in which XXXXXX is A decimal number, which represents the A/D internal code transmitted wirelessly from the scale body. Press [5] key to return the function switch state.

4, electronic hook scale instrument interchange function

When displaying C-Data, press [5] to enter the instrument interchange function. This function has a great impact on the use of electronic scale, so there is password protection. Users should not enter at will.

≡  contacts:Manager Wang ≡  mobile:13930926376 ≡  Telephone:0319-2158243 ≡  fax:0319-2159641 ≡  address:No. 1 in Rongtai science and Technology Incubation Park, south of Xingda street, Xingtai Economic Development Zone
河北超達密封
版權所有:邢臺德港電子有限公司       網站備案/許可證號:       

冀公網安備 13050902000085號

技術支持:邢臺市網商軟件開發有限公司
主站蜘蛛池模板: 国产一区二区三区免费看视频 | 国产aⅴ人| 精品久久久久久无码中文字幕一 | 国产一区二区黄色传媒 | 91精品福利资源 | 国产高潮国产高潮久久久 | 国产精品偷窥熟女精品视频 | 99久久国语露脸精品国产色 | 国产成人亚洲精品无码青青草 | 国产品无码一区二区三区在线 | 国产漂亮白嫩美女在线观看 | 国产精品伦人一久二久三久 | heyzo无码一区综合在线播放 | 国产免费一区二区三区香蕉精 | av无码精品一区二区三区 | 2025亚洲爆乳无码专区 | 国产成人无码一区二区三区在线 | 国产一区二区三精品久久久 | 国产尤物aⅴ在线观看不卡 国产尤物av尤物在线看 | 国产在线精品一区二区不卡麻 | 成人免费一区二区三区视频 | 丰满少妇激情进入高清播放 | 91精品国产高清久 | 97人妻免费专区 | 国产原创高清不卡av在线一区 | 成人午夜视频 | 成人碰碰视频公开老师 | 国产品欧美日产在线 | 2025最新国产精品网站 | 国产一区二区三区观看 | av色欲无码人妻中文字幕 | 精品久久久久久中文字幕专区 | 国产精品无码一区二区久久 | 国产成人午夜福利在线视频 | 国产高清黄色在线 | 精品国产高清久久久久久小说 | 国产精品一区二区久久不卡 | 国产无码精品在 | 极品少妇一区二区三区四区视频 | 91精品无码中文字幕在线 | 成人自慰女黄网站免费大全 |